喷雾参数对雾滴沉积性能影响研究
作者:
作者单位:

作者简介:

通讯作者:

中图分类号:

基金项目:

国家重点研发计划项目(2017YFD0700905)


Effects of Spraying Parameters on Droplet Deposition Performance
Author:
Affiliation:

Fund Project:

  • 摘要
  • |
  • 图/表
  • |
  • 访问统计
  • |
  • 参考文献
  • |
  • 相似文献
  • |
  • 引证文献
  • |
  • 资源附件
  • |
  • 文章评论
    摘要:

    为了研究扇形喷嘴不同喷雾方式下的空间沉积情况,利用自行设计的NJS-1型植保风洞,搭建雾滴粒径测试装置与雾滴沉积分布测试装置。选用LURMARK-04F80型标准扇形喷嘴开展雾滴粒径分布与沉积特性试验,分析了喷雾压力与风速对雾滴粒径的影响,同时研究了不同风速、喷雾压力、雾流角及喷头倾角下雾滴沉积特性,并采用3种不同的计算方法对比了雾滴飘移减少百分比的影响因素。雾滴粒径分布试验结果表明,相同风速下,增大喷雾压力会导致DV0.1、DV0.5和DV0.9都变小,同时ΦVol<100μm变大,雾滴谱宽S变化不大;相同压力下,增大风速导致DV0.1和DV0.5变大,DV0.9变化较小,同时ΦVol<100μm变小,雾滴谱宽S减小。雾滴沉积分布试验结果表明,压力从0.2MPa增加至0.4MPa时,水平喷雾平面上,距离喷头2~3m处雾滴沉积量基本呈增加趋势,竖直喷雾平面上,距离地面0.1~0.2m处雾滴沉积量呈增加趋势;风速从1m/s增加至5m/s时,在水平喷雾平面以及竖直喷雾平面上,雾滴沉积量整体呈增加趋势;雾流角从-15°变化到15°时,在水平喷雾平面以及竖直喷雾平面上,雾滴沉积量明显加大;喷头倾角从0°变化到30°时,在水平喷雾平面以及竖直喷雾平面上,总体趋势是喷头倾角越大,沉积量越低,但差异不大;同时与参考喷雾相比较,采用3种计算方法得到的雾滴飘移减少百分比(DPRP)表明,喷雾压力、风速以及雾流角对雾滴飘移减少百分比影响较大,特别是侧风风速影响尤为显著。该研究可为田间喷雾作业参数的选择提供试验数据指导。

    Abstract:

    In order to research on the effect of different spraying modes of fan nozzle on spatial deposition of droplet, droplet size measurement test device and deposition distribution test device were established by using NJS-1 wind tunnel for plant protection. The droplet size measurement test device was mainly composed of spray system, laser particle size analyzer and so on. The nozzle was mounted on the vertical reciprocating guide rail with the moving speed of 6.7cm/s and the horizontal distance between the nozzle and the laser beam was 30cm. During the tests, the spray pressure was firstly stabilized, and then the laser particle size analyzer was turned on. In order to sample the entire spray stoke area, the spraying nozzle was controlled by a singlechip microcomputer to move at a certain speed. Droplet deposition distribution test device was mainly composed of spray system, wind tunnel system, acquisition system and so on. During the tests, the flow rate of the spray nozzle was controlled by an electronic timer to open/close the solenoid valve to ensure that the spray time of each test was fixed at 10s. The fluorescent tracer BSF was selected as the spray medium and was mixed with water at the ratio of 0.30g/L. After each spray test, the collection line was placed in the plastic bag with 30mL deionized water for full oscillation washing, the amount of fluorescent agent content was determined by the calibrated fluorescence analyzer for each test eluent. The LURMARK-04F80 standard fan nozzle was used in the dropsize distribution and deposition performance tests. The effects of spray pressure and wind speed on droplet size and the influence of wind speed, spray pressure, spray orientation and nozzle direction on droplet deposition were investigated. Three calculation models were employed to compare different influence factors of droplet drift reduction percentage. The results of droplet size distribution experiments showed that at the same wind speed, the increase of spray pressure would cause the decrease of DV0.1, DV0.5 and DV0.9 and the increase of ΦVol<100μm, but minor changes of droplet spectrum width S; under the same spray pressure, the increase of wind speed would cause the increase of DV0.1 and DV0.5 but minor change of DV0.9, the decrease of ΦVol<100μm and droplet spectrum width S was from 1.44 to 1.17. The results of droplet deposition distribution tests showed that when spray pressure was increased from 0.2MPa to 0.4MPa, in the plane parallel to spray direction,the droplet deposition was increased at 2~3m from the spray nozzle,droplet deposition was decreased when spray pressure was increased far away from the nozzle, in the plane vertical to spray direction, the droplet deposition was increased at 0.1~0.2m from the ground and increased when spray pressure was increased, in the middle position, the droplet deposition was decreased when spray pressure was increased, the droplet deposition was close to zero at the height nearest to the nozzle. When wind speed was increased from 1m/s to 5m/s, the droplet deposition was increased on both planes parallel and vertical to spray direction; when the nozzle direction was changed from -15° to 15°, the droplet deposition was increased on both planes parallel and vertical to spray direction; when the nozzle direction was changed from 0° to 30°, the droplet deposition was decreased with the increase of nozzle direction on both planes parallel and vertical to spray direction with minor difference. Compared with the reference spray, the values of DPRP obtained from three calculation models showed that spray pressure, wind speed and spray orientation greatly influenced the droplet drift reduction percentage, especially the crosswind speed. This study can provide experimental data guidance for the selection of spray parameters for spray operation in the field. 

    参考文献
    相似文献
    引证文献
引用本文

丁素明,薛新宇,董祥,顾伟,周晴晴.喷雾参数对雾滴沉积性能影响研究[J].农业机械学报,2020,51(s2):308-315. DING Suming, XUE Xinyu, DONG Xiang, GU Wei, ZHOU Qingqing. Effects of Spraying Parameters on Droplet Deposition Performance[J]. Transactions of the Chinese Society for Agricultural Machinery,2020,51(s2):308-315.

复制
分享
文章指标
  • 点击次数:
  • 下载次数:
  • HTML阅读次数:
  • 引用次数:
历史
  • 收稿日期:2020-07-31
  • 最后修改日期:
  • 录用日期:
  • 在线发布日期: 2020-12-10
  • 出版日期: 2020-12-10